222, 223, 622 and 623 Table Drawer Heater System Troubleshooting
Model |
222-016 | 223-016 | 622-008 | 623 -008,-010, -011 |
Serial Number | All | All | All | All |
This illustration shows only the components that affect the drawer heater system. A detailed description of current flow during this function also appears below.
Drawer Heater Operation
115 VAC is supplied directly to the Drawer Heater switch thru two fuses on the Distribution Board. Voltage bypasses the Main PC Board. When the drawer heater switch is turned ON, current flows to the heater plate. When the heater plate is energized, it warms the contents of the drawer.
Drawer Heater System Troubleshooting
Function | Status | Problem |
---|---|---|
Drawer Heater |
|
Does not heat up. |
Heater Switch |
|
Does not illuminate. |
1.Loose Wire Connections between heater switch and power cord.
2.Distribution Board Fuses. Refer to: Fuses
3.Perform Drawer Heater Switch Test. Refer to: Drawer Heater System
Function | Status | Problem |
---|---|---|
Drawer Heater |
|
Does not heat up. |
Heater Switch |
|
Ok, illuminates. |
1.Loose Wire Connections between heater switch and power cord.
2.Perform Drawer Heater Switch Test. Refer to: Drawer Heater System
Function | Status | Problem |
---|---|---|
Drawer Heater |
|
Ok. |
Heater Switch |
|
Does not illuminate. |
1.Perform Drawer Heater Switch Test. Refer to: Drawer Heater System